Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-25
ZERO/SPAN AVERAGING TIME:
CURRENTLY: 60 SEC
SET TO: 90 SEC ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Zero/Span Ratio screen is used to adjust the ratio of zeros to spans. For
example, if this value is set to 1, a span check will follow every zero check.
If this value is set to 3, there will be two zero checks between each
zero/span check. This value may be set from 1 to 10, with 1 as default.
●
In the Main Menu, choose Calibration > Zero/Span Check >
Zero/Span Ratio
.
ZERO/SPAN RATIO:
CURRENTLY: 1:1
SET TO: 2:1 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Instrument Controls menu contains a number of items. The software
controls listed in this menu enable control of the listed instrument
functions.
●
In the Main Menu, choose
Instrument Controls
.
INSTRUMENT CONTROLS:
>OZONATOR
PMT SUPPLY
AUTO/MANUAL MODE
DATALOGGING SETTINGS
COMMUNICATION SETTINGS
I/O CONFIGURATION
TEMPERATURE COMPENSATION
ª
RANGE AVG DIAGS ALARM
PRESSURE COMPENSATION
SCREEN CONTRAST
SERVICE MODE
DATE/TIME
Zero/Span Ratio
Instrument Controls
Menu