Operation
Diagnostics Menu
3-62
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-
scale voltage, zero sets the analog outputs to 0 volts, and normal operation.
The example below shows the selected output state “ALL” is set to normal.
●
In the Main Menu, choose Diagnostics > Test Analog Outputs >
ALL,
Voltage Channel 1-6
,
or
Current Channel 1-6
.
SET ANALOG OUTPUTS:
SETTING: ALL
OUTPUT SET TO: NORMAL
©
SET TO FULL SCALE
ª
SET TO ZERO
§
RESET TO NORMAL
RANGE AVG DIAGS ALARM
The Instrument Configuration screen displays information on the
hardware configuration of the instrument.
Note
If the analyzer is in service mode, pressing
on the item will
toggle it yes or no (with the exception of purchased options such as
dilution and auto calibration).
▲
●
In the Main Menu, choose Diagnostics >
Instrument Configuration
.
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
ZERO/SPAN VALVES YES
LAG VOLUME NO
PERM DRYER NO
CONVERTER MOLY
DILUTION NO
AUTO CALIBRATION NO
RANGE AVG DIAGS ALARM
The Contact Information screen displays the customer service information.
●
In the Main Menu, choose Diagnostics >
Contact Information
.
Set Analog Outputs
Instrument
Configuration
Contact Information