Operation
Service Menu
3-80
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
The Automatic Input Calibration screen is used to do an automatic
calibration of the input board A/D stages. A message will be displayed after
the optimum setting has been determined.
Note
The measurement system and the PMT are both shut off inside this
screen.
▲
●
In the Main Menu, choose Service > Input Board Calibration >
Automatic Input Calibration
.
●
Press
to leave warning screen and begin automatic calibration.
The Input Frequency Display screen is used to manually adjust the input
board gain. This may be used as a troubleshooting tool for the input board.
The gain setting and test mode are reset upon exiting this screen.
●
In the Main Menu, choose Service > Input Board Calibration >
Input
Frequency Display
.
●
Press
to leave warning screen.
●
Use
and
to toggle the test signal and bypass the PMT.
●
Use
and
to change the gain between 1, 10 and 100.
Automatic Input Calibration
Input Frequency Display
INPUT BOARD CALIBRATION:
TO CALIBRATE
** WARNING **
x
THIS ACTION MAY REQUIRE
xx
xx
RECALIBRATION OF THE
XXXX
ENTIRE MEASUREMENT SYSTEM!
RANGE AVG DIAGS ALARM
GAIN 1 - MANUAL:
FREQ = 4500
D/A = N/A ?
ARE YOU SURE YOU WANT TO?
x
§¨
CHG GAIN
SAVE VALUES
RANGE AVG DIAGS ALARM
INPUT BOARD CALIBRATION:
TO CALIBRATE
** WARNING **
x
THIS ACTION MAY REQUIRE
xx
xx
RECALIBRATION OF THE
XXXX
ENTIRE MEASUREMENT SYSTEM!
RANGE AVG DIAGS ALARM
INPUT BOARD CALIBRATION:
** IN PROGRESS **
D/A = 2047
TARGET = 0
FREQ = 0 0
ZERO = 0 CNT = 1
RANGE AVG DIAGS ALARM