Operation
Service Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-79
The Input Board Calibration menu is used to initiate a calibration of the
input A/D stages. The input board calibration menu is visible only when
the instrument is in service mode. For more information on the service
mode, see “
Service Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Input Board Calibration
.
INPUT BOARD CALIBRATION:
>MANUAL INPUT CAL
AUTOMATIC INPUT CAL
INPUT FREQUENCY DISP
RANGE AVG DIAGS ALARM
The Manual Input Calibration screen is used to do a manual calibration of
the input board A/D stages per the following procedure:
Note
The measurement system and the PMT are both shut off inside this
screen.
▲
1.
In the Main Menu, choose Service > Input Board Calibration >
Manual Input Calibration
.
2.
Press
to leave warning screen.
3.
Make a note of the frequency at gain of 1.
4.
Use
and
to change the gain between 10 and 100.
5.
Use
and
to increment or decrement the D/A counts so
the frequency at gain 100 is equal to the frequency at gain 1.
6.
Press
to save new input board calibration.
Input Board Calibration
Manual Input Calibration