Servicing
Front Panel Board Replacement
7-48
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
9.
At the Flow Sensor Cal menu, press
to select Span.
The Calibrate Flow Span screen appears.
10.
Wait at least 10 seconds for the reading to stabilize, use
and
to enter the flow sensor reading, and press
to
save the value.
11.
Install the cover.
Use the following procedure to replace the front panel board (
Figure 7–
23
).
Equipment Required:
Front panel board
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
2.
Remove the three ribbon cables and the two-wire connector from the
front panel board.
3.
Pop off the board from the two top mounting studs and remove the
board by lifting it up and off the slotted bottom support.
4.
Replace the front panel board by following previous steps in reverse.
Front Panel Board
Replacement