Operation
Instrument Controls Menu
3-36
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
COMMUNICATION SETTINGS:
>BAUD RATE
INSTRUMENT ID
COMMUNICATION PROTOCOL
STREAMING DATA CONFIG
RS-232/RS-485 SELECTION
TCP/IP SETTINGS
RANGE AVG DIAGS ALARM
The Baud Rate screen is used to set the RS-232/RS-485 interface baud rate.
Baud rates of 1200, 2400, 4800, and 9600, 19200, 38400, 57600, and
115200 are available.
●
In the Main Menu, choose Instrument Controls > Communication
Settings >
Baud Rate
.
BAUD RATE:
CURRENTLY: 9600
SET TO: 19200 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The Instrument ID screen allows the operator to edit the instrument ID.
The ID is used to identify the instrument when using the C-Link or
MODBUS protocols to control the instrument or collect data. It may be
necessary to edit the ID number if two or more of the same instrument are
connected to one computer. Valid Instrument ID numbers are from 0 to
127. The Model 42
i
Trace Level has a default Instrument ID of 42. For
more information about the Instrument ID, see Appendix B “
C-Link
Protocol Commands
” or Appendix C “
MODBUS Protocol
”.
●
In the Main Menu, choose Instrument Controls > Communication
Settings >
Instrument ID
.
Baud Rate
Instrument ID