C-Link Protocol Commands
Datalogging
B-20
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
10:11 04-06-05 flags DD008000 no 0.000 nox 0.000 no2
0.000 1 lono 147.500 lonox 0.000 lono2 0.000 1 pmtv
805.491 tempal 1 pres 172.278 pcal 150.000 smplf 0.000
ozonf 0.050 hiavgtime 10 loavgtime 10 nobkg 0.000
noxbkg 0.000 nocoef 1.000 noxcoef 1.000 no2coef 1.000
lonocoef 1.000 lonoxcoef 1.000 lono2coef 1.000 norange
100000.000 noxrange 100000.000 no2range 100000.000
lonorange 100000.000 lonoxrange 100000.000 lono2range
100000.000
lrec
srec
lrec
xxxx yy
srec
xxxx yy
lrec
aa:bb oo-pp-qq yy
srec
aa:bb oo-pp-qq yy
xxxx
= the number of past records
yy
= the number of records to return (1 to 10)
aa
= hours (01 to 24)
bb
= minutes (01 to 59)
oo
= month (01 to 12)
pp
= day (01 to 31)
= year
These commands output long or short records and dynamic data. The
output format is determined by the “set lrec format”, and “set srec format”
commands. The logging time is determined by the “set lrec per” and “set
srec per” commands. In dual range, the long records and short records
contain the high and low NO and NO
x
concentrations. In single range the
low NO and low NOx values are set to 0 and the high NO and high NO
x
are used. In NO or NO
x
only mode, the pertinent high value used, other
concentrations are set to 0. Concentrations are stored in either ppb or
μg/m
3
.
In the following example, there are 740 long records currently stored in
memory. When the command lrec 100 5 is sent, the instrument counts
back 100 records from the last record collected (record 740), and then
returns 5 records: 640, 641, 642, 643, and 644. For details on how to
decode the flag fields within these records, see the “flags” command.