Operation
Calibration Menu
3-22
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
The Calibrate NO Coefficient screen is used to adjust the NO span
concentration while sampling span gas of known concentration. All
calibration screens operate the same way. Therefore, the following
description of the NO calibration screen applies to the NO
2
and NO
x
calibration screens as well.
The display shows the current NO concentration reading and the current
NO range. The next line of the display is where the NO calibration gas
concentration is entered.
It is important to note the averaging time when calibrating. The longer the
averaging time, the more accurate the calibration will be. To be most
accurate, use the 300-second averaging time. For more information about
calibration, see Chapter 4, “
Calibration
”.
●
In the Main Menu, choose Calibration >
Cal NO, NO2 or NOx
Coefficient
.
CALIBRATE NO:
NO: 25.5
SPAN CONC: 0001.00
0
?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Calibrate Pressure screen is used to set the calibration pressure.
The Calibration pressure should be set only when calibrating the
instrument.
●
In the Main Menu, choose Calibration >
Calibrate Pressure
.
CALIBRATE PRESSURE:
CUR CAL PRES: 150.0 mmHg
CURRENT PRES: 232.5 mmHg ?
SAVE CURRENT PRESSURE
AS CAL PRESSURE
RANGE AVG DIAGS ALARM
Calibrate NO, NO
2
or
NO
x
Coefficients
Calibrate Pressure