Installation
Setup Procedure
2-2
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Remove Packing
(2 pieces)
Figure 2–1.
Remove the Packing Material
4.
Check for possible damage during shipment.
5.
Check that all connectors and circuit boards are firmly attached.
6.
Re-install the cover.
7.
Remove any protective plastic material from the case exterior.
8.
Remove the external pump from its shipping container and place next
to the instrument.
Use the following procedure to setup the instrument:
1.
Connect the sample line to the SAMPLE
bulkhead on the rear panel
(
Figure 2–2
). Ensure that the sample line is not contaminated by dirty,
wet, or incompatible materials. All tubing should be constructed of
FEP Teflon®, 316 stainless steel, borosilicate glass, or similar tubing
with an OD of 1/4-inch and a minimum ID of 1/8-inch. The length of
the tubing should be less than 10 feet.
Setup Procedure
Remove Packing
(2 pieces)
Units without Optional I/O Board
Units with Optional I/O Board