C-Link Protocol Commands
Hardware Configuration
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
B-35
Send:
contrast
Receive: contrast
5:50%
set contrast
level
This command sets the screen’s
level
of contrast, according to
Table B–7
.
The example below sets the contrast level to 50%.
Send:
set
contrast
5
Receive:
set contrast 5 ok
Table B–7.
Contrast Levels
Level
Contrast Level
0 0%
1 10%
2 20%
3 30%
4 40%
5 50%
6 60%
7 70%
8 80%
9 90%
10 100%
conv set temp
This command reports the temperature that the NO
2
converter is set to.
The example below reports that the converter temperature is set to 325 °C.
Send:
conv set temp
Receive:
conv set temp 325 deg C
set conv set temp
value
This command sets the temperature that the NO
2
converter is set to, where
value
is an integer representing dwgrees C. The example below sets the
converter temperature to 325 °C.
Send:
set conv set temp
Receive:
set conv set temp 325 deg C ok
date
This command reports the current date. The example below reports the
date as December 1, 2004.