Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-51
streaming data to report what data is being sent out. The descriptor may be
from 1 to 3 characters in length, and defaults to IN1 to IN8 (user input
channel number).
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Descriptor
.
ANALOG INPUT DESCRIPTOR:
CURRENTLY: IN1
I
N1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The Units screen allows the user to enter the units for the selected analog
input channel. The units are displayed on the diagnostic screen and in
datalogging and streaming data. The units may be from 1 to 3 characters in
length, and defaults to V (volts).
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Units
.
ANALOG INPUT UNITS:
CURRENTLY: V
V
N1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal, from 0 to 6, with a default of 2.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Input Config > Select Channel >
Decimal Places
.
Units
Decimal Places