Operation
Service Menu
3-82
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
ª
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog output. The operator must connect a meter
to the output and adjust the output until it reads 0.0 V on the meter.
●
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel >
Calibrate Zero
.
ANALOG OUTPUT CAL: ZERO
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 100
SAVE VALUE
©ª
INC/DEC
SET OUTPUT TO: 0.0 V
RANGE AVG DIAGS ALARM
The Analog Output Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog output. The operator must connect
a meter to the output and adjust output until it reads the value shown in
the set output to: field.
●
In the Main Menu, choose Service > Analog Out Calibration > Select
Channel >
Calibrate Full Scale
.
Analog Output Calibrate
Zero
Analog Output Calibrate
Full-Scale