Servicing
Solenoid Valve Replacement
7-24
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Refer to “
Removing the Measurement Bench and Lowering the
Partition Panel
” in this chapter to lower the partition panel, then
proceed to the next step below.
2.
Disconnect solenoid from the Measurement Interface board. Note
electrical connections to facilitate re-connection.
3.
Remove plumbing from solenoid. Note plumbing connections to
facilitate re-connection.
4.
Pull solenoid valve from mounting clip.
5.
To replace solenoid, follow previous steps in reverse.
6.
Re-install the measurement bench. Refer to “
Removing the
Measurement Bench and Lowering the Partition Panel
” in this chapter.
Figure 7–12.
Replacing the Solenoid Valve