Operation
Alarms Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-71
NO CONCENTRATION:
ACTUAL MIN: 0.0
SET MIN TO: 000
2
0.00 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Minimum Trigger screen allows the user to view and set the NO,
NO
2
, and NO
x
concentration alarm trigger type to either floor or ceiling.
The minimum alarm may be programmed as a floor trigger (alarm is
triggered when the concentration falls below the minimum value) or a
ceiling trigger (alarm is triggered when the concentration goes above the
minimum value).
●
In the Main Menu, choose Alarms > Select Concentration >
Min
Trigger
.
MIN TRIG(CEILING/FLOOR):
ACTUAL TRIGGER: CEILING
SET TRIGGER TO: FLOOR ?
TOGGLE AND SAVE VALUE
RANGE AVG DIAGS ALARM
The Zero Concentration screen displays the current zero concentration and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 0 to 2 ppb and 0 to 5 μg/m
3
. The minimum alarm may be
programmed as a floor trigger (alarm is triggered when the concentration
falls below the minimum value) or a ceiling trigger (alarm is triggered when
the concentration goes above the minimum value). If the zero
concentration goes beyond either the minimum or maximum limit, an
alarm is activated. The word “ALARM” appears in the Run screen and in
the Main Menu.
●
In the Main Menu, choose Alarms >
Zero Concentration
.
Min Trigger
Zero Concentration