Operation
Alarms Menu
3-70
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
auto calibration screens are visible only if the auto calibration option is
enabled and function the same way.
●
In the Main Menu, choose Alarms >
Zero
or
Span Autocal
.
ZERO AUTO CALIBRATION:
ALARM: OK
RESPONSE: 5.0
RANGE AVG DIAGS ALARM
The NO Concentration screen displays the current NO concentration and
sets the minimum and maximum alarm limits. Acceptable alarm limits
range from 0 to 200 ppb and 0 to 500 μg/m
3
. The minimum alarm may be
programmed as a floor trigger (alarm is triggered when the concentration
falls below the minimum value) or a ceiling trigger (alarm is triggered when
the concentration goes above the minimum value). If the NO
concentration goes beyond either the minimum or maximum limit, an
alarm is activated. The word “ALARM” appears in the Run screen and in
the Main Menu. The NO
2
and NO
x
concentration screens function the
same way.
●
In the Main Menu, choose Alarms >
NO, NO2, or NOx
Concentration
.
NO CONCENTRATION:
ACTUAL 62.7
>MIN 0.0
MAX 200.0
MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
The Minimum NO Concentration alarm limit screen is used to change the
minimum NO concentration alarm limits. The minimum and maximum
NO, NO
2
, and NO
x
concentration alarm limit screens function the same
way.
●
In the Main Menu, choose Alarms > Select Concentration >
Min
or
Max
.
NO, NO
2
, and NO
x
Concentration
Min and Max NO, NO
2
, and NO
x
Concentration Limits