C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
B-45
D to A
Function
Single Range
Dual Range
Auto Range
11
Current Output
Not Used
Low NO
2
Not
Used
12
Current Output
Not Used
Low NO
x
Not
Used
list var aout
list var dout
list var din
These commands report the list of index numbers, and the variables
(associated with that index number) available for selection in the current
mode (determined by single/dual/auto, gas mode) for analog output, digital
output and digital inputs. The index number is used to insert the variable
in a field location in a list using “set sp
field index
”. The example below
reports the list of analog output, index numbers, and variables.
Send:
list var aout
Receive:
list var aout
index
variable
0
none
1
no
2
no2
3
nox
18
intt
19
rctt
20
pmtt
21
convt
25
pres
26
smplf
27
pmtv
28
ain1
29
ain2
30
ain3
31
ain4
32
ain5
33
ain6
34
ain7
35
ain8
relay stat
This command reports the current relay logic normally “open” or normally
“closed,” if all the relays are set to same state, that is all open or all closed.
The example below shows that the status when all the relays logic is set to
normally “open”.
Send:
relay
stat
Receive:
relay stat open