Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 42
i
Trace Level Instruction Manual
3-45
The Non-Alarm status screen allows the user to select the non-alarm status
for the selected relay output. The selected item is shown by “<--” after it.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Output Relay Settings > Select Relay > Instrument State >
Non-Alarm
.
NON ALARM STATUS ITEMS:
>NONE
AUTORANGE (NOx)
SERVICE
UNITS
ZERO MODE
SPAN MODE
NO MODE
ª
RANGE AVG DIAGS ALARM
NOX MODE
SAMPLE MODE
PURGE MODE
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note
The digital inputs must be asserted for at least one second for the
action to be activated.
▲
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Setting
s
.
DIGITAL INPUT SETTINGS:
>1 NOP NO MODE
2 NOP NOX MODE
3 NOP SET BACKGROUND
4 NOP CAL TO LO SPAN
5 NOP AOUTS TO ZERO
6 NOP AOUTS TO FS
7 NOP NONE
ª
RANGE AVG DIAGS ALARM
The Logic State screen is used to change the I/O relay to either normally
open or normally closed. The default state is open, which indicates that a
relay connected between the digital input pin and ground is normally open
Non-Alarm
Digital Input Settings
Logic State