Operation
Alarms Menu
3-72
Model 42
i
Trace Level Instruction Manual
Thermo Fisher Scientific
ZERO CONCENTRATION:
ACTUAL 0.09
>MIN 0.00
MAX 2.00
MIN TRIGGER CEILING
RANGE AVG DIAGS ALARM
The Minimum NO Concentration alarm limit screen is used to change the
minimum NO concentration alarm limits. The minimum and maximum
NO, NO
2
, and NO
x
concentration alarm limit screens function the same
way.
●
In the Main Menu, choose Alarms > Zero Concentration >
Min
or
Max
.
NO CONCENTRATION:
ACTUAL MIN: 0.0
SET MIN TO: 000
2
0.00 ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Minimum Trigger screen allows the user to view and set the Zero
concentration alarm trigger type to either floor or ceiling. The minimum
alarm may be programmed as a floor trigger (alarm is triggered when the
concentration falls below the minimum value) or a ceiling trigger (alarm is
triggered when the concentration goes above the minimum value).
●
In the Main Menu, choose Alarms > Zero Concentration >
MinTrigger
.
MIN TRIG(CEILING/FLOOR):
ACTUAL TRIGGER: CEILING
SET TRIGGER TO: FLOOR ?
TOGGLE AND SAVE VALUE
RANGE AVG DIAGS ALARM
Min and Max Zero Concentration
Limits
Min Trigger