Recommended minimal washer /disinfector parameters:
Temperature
Time
Heated Wash
60°C (140°F)
2 Min.
Heated Tap Water Rinse
60°C (140°F)
20 Sec.
Heated Deionized Water Rinse
82°C (180°F)
1 Min.
Forced Air Drying
116°C (240°F)
9 Min.
10.3 Sterilization Instructions
NOTE:
1. For components with moving parts, lubricate joints with a
steam-permeable, water soluble component lubricant prior
to sterilization.
2. Components should be sterilized in the opened or unlocked
position. Central knob of the articulating arms must be
opened for sterilization.
3. Components should be sterilized by standard cycles using
steam with the procedures listed below.
Preparation for Sterilization:
Note: Use only FDA Cleared CSR Sterilization Wrap for
component packaging.
• Each component should be wrapped in two layers of FDA
Cleared CSR Wrap using the simultaneous double-
wrapping equal fold technique.
• Chemical Indicator Tape should be used to secure
packaging and for labeling the contents. Note: Chemical
Indicator Tape will change color or display diagonal stripes
when exposed to temperatures of 121°C (250°F).
Note: Only use wrappers validated for use in double simultaneous
wrapping.
Two single-layer wrappers or one bonded double wrapper can
be used. Simultaneous double wrapping refers to when two
layers of wrap material are wrapped together simultaneously
using the fold technique described below:
1. Place the edge of the wrapper parallel with the edge of
the table. Place the device in the center of the wrapper
parallel with the edge of the wrapper.
2. Fold the edge of the wrapper over the top of the contents,
covering the entire item. Fold the edge back over itself
(toward the technician) to form a cuff. This will facilitate
aseptic opening of the pack when used.
Содержание HH1001 Main Arm
Страница 202: ... 2020 IFU 0100 rev3 202 1 定義 記号説明 参照番号 バッチコード 製品数量 欧州医療機器規則に準拠 製造年月日 製造業者 欧州共同体域内の販売責任者 未滅菌で供給される製品 製品が医療機器であることを示す ...
Страница 212: ... 2020 IFU 0100 rev3 212 ...
Страница 225: ... 2020 IFU 0100 rev3 225 800 624 4422 수신자 부담 전화 미국 전용 또는 이메일 EndoCustomerSupport stryker com ...
Страница 228: ... 2020 IFU 0100 rev3 228 1 定义 符号说明 产品编号 批号 产品数量 符合欧盟医疗器械法规 制造日期 制造商 欧盟授权代表 非无菌产品 表示产品为医疗器械 ...
Страница 238: ... 2020 IFU 0100 rev3 238 1 定義 符號說明 參考號碼 批號 產品數量 符合歐洲醫療器材法規 製造日期 製造商 歐盟授權代表 產品出貨時並未滅菌 所示產品為一醫療器材 ...