Flow-X─ Function Reference
4-83
Flow-X General functions - fxAPI_Dens15C_NGL_LPG
Function outputs
Name
Remark
EU
SW tag
Alarm
Fallb
ack
Status
0: Normal
1: Input argument out of range
2: Calculation error
3: No convergence
-
STS
FIOOR
CALCERR
NOCONV
1
Output density
Depending on the conversion
method this represents the density
either at 15
°
C and the equilibrium
pressure or the observed
temperature and pressure
kg/m3
DENS
0
CTL
Volume correction factor for
temperature.
Value will be rounded according to
input 'API 11.2.4 rounding'
-
CTL
1
CPL
Volume correction factor for
pressure
Value will be rounded according to
input 'API 11.2.2M rounding'
-
CPL
1
CTPL
Combined volume correction factor
CTPL = CTL * CPL
-
CTPL
1
F
Compressibility factor
The output value will be either
rounded or not depending input
'API rounding'
1/bar
F
0
Equilibrium
pressure
The equilibrium pressure calculated
by GPA TP-15
Will be set to 0 when equilibrium
pressure is below atmospheric
pressure
bar(a)
EQUIPC
UR
0
CTL calc out of
range
With respect to the API 11.2.4
standard the combination of input
values is:
0: In Range
1: Out of Range
The following range checks apply:
Conversion method 1: observed ->
standard
0.21 <= RD <= 0.74
with RD = Input density /
999.016/CPL
-46 <= T <= 93 °C
Table 23E reference fluid
ranges
Conversion method 1: standard ->
observed
351.7 <= Input density <=
687.8 kg/m3
-46 <= T <= 93 °C
Table 23E reference fluid
ranges
CTLOOR
0
Содержание flow-x
Страница 1: ...Function Reference Certified flow calculations Flow and batch calculations Worksheet functions...
Страница 331: ...Flow X Function Reference 5 331 Flow X IO Functions fxPT100Table...
Страница 349: ...Flow X Function Reference 6 349 This page is intentionally left blank for user notes...
Страница 350: ...6 350 Flow X Function Reference This page is intentionally left blank for user notes...