4-270
Flow-X─ Function Reference
Flow-X General functions - fxSarasota_C
correction is enabled
d
0
Constant from
calibration certificate
Note: value required in
lbm/ft3
lbm/ft3
D0
0..2000
0
0
Constant from
calibration certificate
s
T0
0..1e6
0
K
Spool calibration
constant from
calibration certificate
-
K
0..3e3
0
Temperature
coefficient
Constant from
calibration certificate
s/°F
TEMPCOEF
-1e6..1e6
0
Pressure coefficient
Constant from
calibration certificate
s/psi
PRESCOEF
-1e6..1e6
0
Function
outputs
Remark
EU
SW tag
Alarm
Fallback
Status
0: Normal
1: Input argument out of range
2: Calculation error
STS
FIOOR
CALC
Corrected
density
Density corrected for
temperature and pressure
lbm/ft3
CORDENS
0
Calculations
The corrected density ρ
C
is calculated by
C
C
C
C
C
K
d
2
0
CAL
COEF
CAL
COEF
C
p
p
p
t
t
t
0
Where:
ρ
C
The corrected density
lbm/ft
3
d
0
Obtained from the calibration certificate
lbm/ft
3
0
Obtained from the calibration certificate
s
K
Obtained from the calibration certificate
-
p
COE
F
Obtained from the calibration certificate
s/psi
t
COE
F
Obtained from the calibration certificate
s/°F
t
Line temperature
°F
t
CAL
Reference temperature
°F
p
Line temperature
psig
p
CAL
Reference pressure
psig
C
Time periodic input corrected for temperature and
pressure
s
τ
Measured time period
s
Содержание flow-x
Страница 1: ...Function Reference Certified flow calculations Flow and batch calculations Worksheet functions...
Страница 331: ...Flow X Function Reference 5 331 Flow X IO Functions fxPT100Table...
Страница 349: ...Flow X Function Reference 6 349 This page is intentionally left blank for user notes...
Страница 350: ...6 350 Flow X Function Reference This page is intentionally left blank for user notes...