Conductivity
ProLab 4000
14
8
ba75646d03
12/2012
10.2.7
Selecting and adjusting the nonlinear
temperature compensation
TC nLin1
The nonlinear temperature compensation is carried out according to
the following equation:
The equation includes two temperature coefficients, TC
1
and TC
2
. In
the ProLab 4000 meter, the temperature coefficients TC
1
and TC
2
are
stored for the following test samples:
Suitable temperature coefficients TC
1
and TC
2
for solutions that are not
mentioned here can be taken from tables found in further reading and
then entered in the meter.
X
Tref
= X
T
/ (1 + TC
1
·[T-TRef] + TC
2
·[T-TRef]
2
)
X
T
= Conductivity at temperature T
[S/cm]
X
Tref
= Conductivity at reference temperature Tref
[S/cm]
TC
1
= Temperature
coefficient
1
[1/°C]
TC
2
= Temperature
coefficient
2
[1/°C
2
]
T
= Temperature
[°C]
Tref
= Reference temperature
[°C]
HCl
NaOH
NaCl
KCl
Concentration c
[mol/l]
1,0
0,67
0,85
0,67
Temperature range
[°C]
0 … 55
0 … 40
0 … 40
0 … 40
Tref [°C]
25
25
25
25
1
Highlight the channel of the conductivity measuring cell in the
measured value display.
2
If necessary, select the measured parameter with
<MODE>
.
3
Open the input window for the temperature coefficients with
Calibration
/
TC calibration
/
TC nLin1...
.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......