ORP voltage
ProLab 4000
92
ba75646e03
12/2012
Criteria
With identical measurement conditions, the following applies:
4
With
<OK>
activate the
Stability control
function.
The background color is red as long as the measured value
does not meet the stability criterion.
As soon as the measured value is stable, the background color
is green. The current measurement data is downloaded to the
interface. The measured values of all channels meeting the cri-
terion for stability control are marked by SC.
Note
When you measure the electromotive force, please note that it may
take some time until the measured value is stable.
Note
You can terminate prematurely the
Stability control
function with
<OK>
manually at any time. If the
Stability control
function is termi-
nated prematurely, the current measurement data is not downloaded
to the interface.
5
Using
<SC>
or
<MODE>
, release the frozen measured value.
The [SC] status display disappears.
Measured param-
eter
Reproducibility
Response time
ORP voltage
better than 0.3 mV
> 15 seconds
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......