Conductivity
ProLab 4000
144
ba75646d03
12/2012
10.2.4
Determining/setting the
temperature compensation
The conductivity of aqueous solutions is mainly caused by the temper-
ature and the concentration of dissolved substances.
Reference temperature
To compare measured values that were determined at different tem-
peratures they have to be converted to a reference temperature. Thus,
a conductivity specification includes a reference temperature specifica-
tion as a matter of principle.
Temperature compensa-
tion
The temperature compensation is used to calculate the conductivity at
the reference temperature from a conductivity value that was measured
at any temperature.
To do so, the temperature dependency (linear or nonlinear) of the con-
ductivity for the solution must be known.
The characteristics of the test sample indicate which procedure is suit-
able for temperature compensation.
For diluted test samples, linear and nonlinear temperature dependen-
cies are stored in the meter.
For other test samples the ProLab 4000 provides the option of deter-
mining the temperature dependency and, if necessary, the concentra-
tion dependency of the conductivity as well.
Characteristics of the test samples
Procedure for temperature
compensation
diluted
wide temperature range
nonlinear (
TC nLF
)
diluted
small temperature range
linear (
TC lin
)
concentrated
wide temperature range
nonlinear (
TC nLin1
...2)
wide concentration range
wide temperature range
nonlinear (
TC nLin3
,
TC nLin4
)
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......