ProLab 4000
What to do if...
197
ba75646e03
12/2012
16
What to do if...
16.1
pH and ORP measurement
Error message
Err1
Error message
Err2
Error message
Err4
Error message,
Error
Cause
Remedy
pH electrode:
– Measured value outside the
measuring range
– Use a suitable electrode
– Air bubble in front of the dia-
phragm
– Remove air bubble
– Air in the diaphragm
– Extract air or moisten junc-
tion
– Cable broken
– Replace the electrode
– Gel electrolyte dried out
– Replace the electrode
Cause
Remedy
– No electrode connected
– Connect the electrode
– Setting time during calibration
too long
– Adjust temperature if neces-
sary
– Recalibrate
Cause
Remedy
– Temperature not stable during
calibration.
– Adjust temperature if neces-
sary
– Recalibrate
Cause
Remedy
pH electrode:
– The values determined for zero
point and slope of the electrode
are outside the allowed limits.
– Recalibrate
– Diaphragm contaminated
– Clean diaphragm
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......