pH
ProLab 4000
76
ba75646e03
12/2012
* Brand names or trade names are trademarks of their respective owners
protected by law (see page 215).
Calibration points
Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The measuring instrument
determines the following values and calculates the calibration line as
follows:
19
Reagecon TEC
*
2,000
4,010
7,000
10,000
25 °C
20
Reagecon 20
*
2,000
4,000
7,000
10,000
13,000
20°C
21
Reagecon 25
*
2,000
4,000
7,000
10,000
13,000
25 °C
22
Riedel-de Haen
*
2,000
4,000
7,000
10,000
20°C
No.
Buffer set *
pH values
at
Note
The buffers are selected in the
Calibration
sensor menu (see
Displayed calibration data
1-point
Zero point
Slope = Nernst slope (-59.16 mV/pH at 25 °C)
2-point
Zero point
Slope
3-point to
5-point
Zero point
Slope
The calibration line is calculated by linear regression.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......