ProLab 4000
Ion concentration
107
ba75646e03
12/2012
3
When measuring without temperature sensor:
Measure the temperature of the standard solution using a
thermometer.
Open the setting window for the temperature value with
[Man. temp.]
.
Enter the temperature and confirm with
[OK]
.
4
Thoroughly rinse the electrode with distilled water.
5
Immerse the electrode in standard solution 1.
6
Open the
ISE standard input
window with
[Continue]
.
7
Select a value from the
Concentration
list.
or
Enter a value with
<0...9>
.
or
Cancel the calibration with
[Cancel]
.
Note
To enter a concentration in exponential representation such as
1.00 E2 mg/l:
Enter the mantissa with
<0...9>
(here: 1.00).
Enter the exponential sign (E) with
<- .><- .>
.
Enter the exponent with
<0...9>
(here: 2).
8
Start the measurement with
[OK]
.
The measured value is checked for stability (
Stability control
).
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......