Determining the cell constant (calibration)
Proceed as follows to determine the cell constant:
1
Connect a conductivity measuring cell to the meter.
2
Select the channel of the measuring cell in the measured value
display.
3
If necessary, select the parameter, conductivity with
<MODE>
.
4
Start the calibration with
<CAL>
.
The calibration display appears.
5
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
6
Start the measurement with
[OK]
.
The measured value is checked for stability (
Stability control
).
7
Wait for the end of the measurement with Stability control or
accept the calibration value with
[OK]
.
The cell constant determined is displayed.
8
Accept the new calibration values with
[OK]
.
The calibration record is displayed and output to the interface.
The calibration procedure is finished.
or
To switch to the measured value display without accepting the
new calibration values, press
<ESC>
.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......