Temperature compensation
Name of a function that considers the temperature influence on the
measurement and converts it accordingly. Depending on the mea-
sured parameter to be determined, the temperature compensation
functions in different ways. For conductimetric measurements, the
measured value is converted to a defined reference temperature. For
potentiometric measurements, the slope value is adjusted to the
temperature of the test sample but the measured value is not con-
verted.
Temperature function
Name of a mathematical function expressing the temperature behav-
ior of a test sample, a sensor or part of a sensor.
Test sample
Designation of the test sample ready to be measured. Normally, a
test sample is made by processing the original sample. The test
sample and original sample are identical if the test sample was not
processed.
Zero point
Designation for the offset potential of a pH electrode. It is the mea-
surable potential of a symmetrical electrode, the membrane of which
is immersed in a solution with the pH of the nominal electrode zero
point (pH = 7).
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......