ProLab 4000
ORP voltage
91
ba75646e03
12/2012
8
.2.2
Measuring
Stability control
The
Stability control
function checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of
measured values. The background color is red until the measured
value is stable.
Regardless of the setting for
Autom. stability control
(see page 63) in
the
System
menu, you can start the
Stability control
function manually
at any time.
Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a test solution.
1
Perform the preparatory activities (see page 69).
2
Highlight the channel of the electrode in the measured value
display.
3
If necessary, select the U display with
<MODE>
.
4
Immerse the ORP electrode in the test sample.
1
Highlight the channel of the ORP electrode.
2
If necessary, select the U or dU parameter with
<MODE>
.
3
Freeze the measured value with
<SC>
.
The display switches to the display of the selected channel as
necessary.
In the status line, [SC] is displayed.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......