ProLab 4000
Memory
169
ba75646e03
12/2012
11.2.2 Managing the calibration data
The calibration data is stored in the current calibration record (menu,
Calibration
/
Calibration record...
) and in the calibration history (menu,
Calibration
/
Calibration history...
and
Memory
/
Calibration memory
).
The current calibration record of ID sensors is stored in the sensor, of
non ID sensors it is stored in the ProLab 4000.
The meter creates a separate calibration history for each ID sensor cal-
ibrated with the ProLab 4000.
For non ID sensors calibrated with the meter, a calibration history is
created for each measured parameter (pH, ISE, Cond) where the cur-
rent calibration data is appended.
Managing functions for all calibration data are available to the adminis-
trator only.
Viewing the calibration
history
The administrator can view the calibration histories of all sensors and
users.
Erasing a
calibration
history
The administrator can delete the calibration histories of all sensors and
users.
The current calibration data of each sensor is stored separately and is
retained even if the calibration history is deleted.
1
Open the menu,
Memory
/
Calibration memory
/
Display...
.
The
Open file
dialog box pops up.
2
If necessary, select a different folder.
3
Select a file ("*.cal") and confirm with
[OK]
.
The
Calibration memory
window with the calibration data pops
up.
4
If necessary, output the calibration data to the interface with
[
Print]
.
1
Open the menu,
Memory
/
Calibration memory
/
Delete...
.
The
Delete file
dialog box pops up.
2
If necessary, select a different folder.
3
Select a file ("*.cal") and confirm with
[OK]
.
A safely query pops up.
4
Confirm the deletion of the file with
[OK]
.
The file is deleted.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......