ProLab 4000
Ion concentration
119
ba75646e03
12/2012
5
Thoroughly rinse the electrode with deionized water.
6
Immerse the electrode in the sample.
7
Start the measurement with
[Continue]
.
After end of the measurement, the
Standard
pops up.
8
In the
Standard
register, enter the value for
Standard concen-
tration
if necessary.
9
Use
[Dosing]
to automatically add to the sample the adjusted
volume of the standard solution (
Standard volume
) (1 vol% of
sample solution).
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (
Standard volume
) (1 vol% of sample solution).