Ion concentration
ProLab 4000
106
ba75646e03
12/2012
9.3.1
Carrying out calibration
Use two to nine standard solutions in ascending or descending order.
Preparatory activities
Perform the following preparatory activities when you want to calibrate:
Calibration
Note
When calibrating within the framework of a method measurement, it is
always a two-point calibration.
1
Connect an ion sensitive electrode to the meter.
2
Highlight the channel of the electrode in the measured value
display.
3
If necessary, select the ISE display with
<MODE>
.
4
If necessary, open the
Measuring
/
ISE settings
/
Ion type
menu
and select the ion type to be measured.
5
Keep the standard solutions ready.
6
If necessary, measure the temperature of the standard solu-
tions with a thermometer.
7
If necessary, change the unit of the measurement result and
calibration standards in the menu,
Measuring
/
Units
.
1
Perform the preparatory activities (see page 106).
2
Start the calibration with
<CAL>
.
The
ISEX calibration
window opens.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......