ProLab 4000
System functions
65
ba75646e03
12/2012
6.10
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
6.10.1 Resetting the sensor settings
The sensor settings are reset in the sensor menu.
In the measured value display, activate the relevant sensor window
[Tab]
and open the menu with
[Menu]
.
pH
The following settings for pH measurements are reset to the default
settings with the
Reset pHMeasuring
function:
Cond
The following settings for conductivity measurements are reset to the
default settings with the
Reset CondCond-Setup
function:
Note
The calibration data is reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Calibration interval...
7 d
Slope in
mV/pH
Measured parameter
pH
pH resolution
High
Zero point
0 mV
Slope
-59.16 mV/pH
Temperature, manual
25 °C
Setting
Default settings
Calibration interval...
150 d
Measured parameter
ϰ
Cell constant
0.475 cm
-1
(
Calibrated cell constant
)
0.475 cm
-1
(
Adjustable cell constant...
)
Temperature coeffi-
cient
TC nLF
Tref
Tref 25
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......