ProLab 4000
Conductivity
145
ba75646d03
12/2012
Temperature compensa-
tion for diluted solutions
For the measurement of aqueous solutions
in a low concentration range (diluted solution)
in a small temperature range (|T - Tref| < 10 K)
with a fixed reference temperature at 20 °C or 25 °C,
the procedures of linear (TC lin) and nonlinear (TC nLF) temperature
compensation are suitable.
The calculation of the temperature compensation is based on the pre-
set reference temperature, 20 °C or 25 °C. It appears on the display as
[TR20]
or
[TR25]
.
Application tips
Select the following temperature compensation given in the table
according to the respective test sample:
Note
The reference temperature and temperature compensation are set in
the
Measuring
menu (see page 142).
Test sample
Temperature compensation
Display
indica-
tor
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water
nLF
according to EN 27 888
nLF
Seawater
(salinity)
Automatic
nLF
according to IOT
Sal
,
nLF
Other diluted solutions
with known tempera-
ture dependency
Linear temperature compensa-
tion;
Temperature coefficient can be
set in the range
0.001 ... 3.000 %/K
lin
Solutions with
unknown temperature
dependency
No temperature compensation
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......