ProLab 4000
Conductivity
155
ba75646d03
12/2012
Note
The measuring method can only be started if
Ref. temp.
,
Start temp.
and
Temp. interval
are selected so that the reference temperature cor-
responds to the nominal temperature of a measuring point:
Ref. temp.
=
Start temp.
+ x*
Temp. interval
5
Enter the start temperature, end temperature and temperature
interval.
Enter the number of standards and confirm with
[Continue]
.
The display shows:
the current conductivity value
the current temperature
information on the nominal temperature of the next measur-
ing point and the temperature difference between the cur-
rent temperature and the nominal temperature
6
Warm or cool the measuring solution (standard 1) to the start
temperature
Start temp.
.
As soon as the temperature of the measuring point is achieved
and a stable value is recognized, the measurement is automat-
ically started and the dataset stored.
After the measurement, a piece of information about the next
measuring point with the relevant nominal temperature is dis-
played and the temperature difference between the current
temperature and the nominal temperature.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......