Technical Data
ProLab 4000
206
ba75646e03
12/2012
17.2
Measuring ranges, resolution, accuracy
17.2.1 Temperature
17.2.2 pH/ORP
Measuring ranges,
resolution
Variable
Measuring range
Resolution
T [°C]
- 35.0 ... + 150.0
0.1
Accuracy
(± 1 digit)
Variable
Accuracy
Temperature of the
test sample
T [°C]
± 0.1
- 35.0 ... + 150.0
Manual
temperature input
Variable
Range
Increment
T
manual
[°C]
-35 ... + 150
1
Measuring ranges,
resolution
Variable
Measuring range
Resolution
pH
- 2.000 ... + 20.000
- 2.00 ... + 20.00
- 2.0 ... + 20.0
0.001
0.01
0.1
U [mV]
- 2200.0 ... + 2200.0
- 2200 ... + 2200
0.1
1
Accuracy
(± 1 digit)
Variable
Accuracy
Temperature of the test
sample
pH
± 0.002
± 0.01
± 0.1
+ 15 °C ... + 35 °C
U [mV]
/ range
- 2200.0 ... + 2200.0
± 0.1
+ 15 °C ... + 35 °C
- 2200 ... + 2200
± 1
+ 15 °C ... + 35 °C
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......