System functions
ProLab 4000
60
ba75646e03
12/2012
6.5
Selecting the channels for measured value display
In the
Window
/
Channels...
you can define which channels are visible
in the measured value display.
Depending on the number of activated measurement data one, two or
three sensor data are displayed at the same time.
6.6
Interfaces for data download
The meter has three interfaces to which data can be downloaded:
RS232: serial printer or PC
USB-B (
USB Device
): PC
USB-A (
USB Host
): USB printer
You can specify to which interfaces data will be downloaded at the
same time.
Menu item
Description
pH/U/ISE Probe 1
Display or hide the sensor data of the pH/U/ISE
1 channel.
pH/U/ISE Probe 2
Display or hide the sensor data of the pH/U/ISE
2 channel.
Cond Probe
Display or hide the Cond sensor window.
1
Open the
System
/
Data output...
menu.
The
Data output
window opens.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......