ProLab 4000
pH
79
ba75646e03
12/2012
7.3.1
Carrying out a calibration procedure
(example:
AutoCal DIN
)
Make sure the
Calibration
/
pH calibration type
/
Buffer sets
setting is
selected correctly and the
AutoCal
calibration type is marked (see
page 73).
Use any one to five buffer solutions of the selected buffer set.
The
DIN
calibration is described below. With other buffer sets, other
nominal buffer values are displayed. Apart from that, the procedure is
identical.
1
Perform the preparatory activities (see page 78).
2
Start the calibration with
<CAL>
.
The
pHX calibration
window opens.
3
If necessary, check whether the displayed buffer set agrees
with the used buffer set.
4
When measuring without temperature sensor:
Measure the temperature of the standard solution using a
thermometer.
Open the setting window for the temperature value with
[Man. temp.]
.
Enter the temperature and confirm with
[OK]
.
5
Thoroughly rinse the electrode with distilled water.
6
Immerse the electrode in buffer solution 1.
7
Start the measurement with
[Continue]
.
The measured value is checked for stability (Stability control).
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......