Conductivity
ProLab 4000
156
ba75646d03
12/2012
10.2.11Switching off the temperature compensation (TC off)
7
Slowly warm or cool the test sample (standard 1) until the end
temperature
End temp.
is achieved.
Each time the nominal temperature for a measuring point is
reached a dataset is stored.
At the end of the series of measurements for the solution (stan-
dard 1), the corresponding measuring record is displayed.
8
Confirm the record with
[Continue]
.
9
Repeat the steps 5 to 8 for all further solutions (standard 2 to x).
The corresponding calibration record with the determined tem-
perature coefficients is displayed at the end of the calibration.
If the calibration is successful, the determined temperature
dependency for the
TC nLin3
method is stored and used for fur-
ther measurements with this method.
10
Confirm the record with
[Continue]
.
[nLin4] is displayed in the status line of the measuring channel.
1
Highlight the channel of the conductivity measuring cell in the
measured value display.
2
If necessary, select the measured parameter with
<MODE>
.
3
Switch off the usage of a temperature compensation with
Mea-
suring
/
Temperature coefficient
/
TC off
.
The temperature compensation is no longer displayed in the
status line.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......