Ion concentration
ProLab 4000
9
8
ba75646e03
12/2012
9.2
Measuring the ion concentration
9.2.1
Preparatory activities
Perform the following preparatory activities when you want to measure:
Temperature sensor
You can measure with or without a temperature sensor. The TP display
indicator appears when a temperature sensor is connected.
1
Connect an ion sensitive electrode to the meter.
2
Highlight the channel of the electrode in the measured value
display.
3
If necessary, select the ISE display with
<MODE>
.
4
Open the
Measuring
/
ISE settings
/
Ion type
menu and select
the ion type to be measured.
5
With manual temperature input:
– Measure the temperature of the test sample using a ther-
mometer.
– Open the
Measuring
/
Setting of man. temp...
menu.
The
Manual temperature
window opens.
– Enter the measured temperature value and confirm it with
[OK]
.
6
Calibrate the meter with the electrode (see page 103).
or
For measurement with a method, start a method.
The calibration is included in the course of the measuring
method (see page 113 ff.).
Note
While no valid calibration is available, e.g. in the delivery condition,
Error
appears in the measured value display.
Note
If you work with an ion selective electrode without temperature sensor,
you can also use the temperature sensor of another electrode (see
page 62). In this case, the indication from which channel the temper-
ature value is taken appears next to the temperature value.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......