What to do if...
ProLab 4000
19
8
ba75646e03
12/2012
No stable measured
value
Obviously incorrect
measured values
– Electrode broken
– Replace the electrode
Buffer solutions:
– Incorrect buffer solutions
– Select the suitable buffer set
for the buffers used
or
– Replace the buffer solutions
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Replace the buffer solutions
Cause
Remedy
Cause
Remedy
pH electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if neces-
sary
Ele test sample:
– Conductivity too low
– Use a suitable electrode
– Temperature too high
– Use a suitable electrode
– Organic liquids
– Use a suitable electrode
Cause
Remedy
pH electrode:
– Not connected
– Connect the electrode
– Cable broken
– Replace the electrode
– pH electrode unsuitable
– Use a suitable electrode
– Temperature difference
between buffer and test sam-
ple too high
– Adjust temperature of buffer
or sample solutions
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......