Conductivity
ProLab 4000
152
ba75646d03
12/2012
10.2.9
Selecting and determining the nonlinear temperature
compensation
TC nLin3
Note
The measuring method can only be started if
Ref. temp.
,
Start temp.
and
Temp. interval
are selected so that the reference temperature cor-
responds to the nominal temperature of a measuring point:
Ref. temp.
=
Start temp.
+ x*
Temp. interval
1
Highlight the channel of the conductivity measuring cell in the
measured value display.
2
If necessary, select the measured parameter with
<MODE>
.
3
Immerse the measuring cell in the test sample.
4
Open the input window for the measurement of the tempera-
ture dependency with
Calibration
/
TC calibration
/
TC nLin3
.
5
Enter the start temperature (
Start temp.
), end temperature
(
End temp.
) and temperature interval (
Temp. interval
).
Enter the number of standards and the corresponding concen-
trations.
The concentration unit entered here is displayed as the unit of
the measured value display after the calibration is completed.
6
Confirm with
[Continue]
.
The display shows:
the current conductivity value
the current temperature
information on the nominal temperature of the next measur-
ing point and the temperature difference between the cur-
rent temperature and the nominal temperature
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......