Ion concentration
ProLab 4000
100
ba75646e03
12/2012
Criteria
The criteria of the stability control affect the reproducibility of the mea-
sured values.
If possible, always use the highest reproducibility criteria.
If the stability criteria are not met with your test sample and electrode,
you can change to lower reproducibility criteria. The following criteria
can be adjusted:
High
: highest
reproducibility
Medium
: medium reproducibility
Low
:
lowest reproducibility
Note
With increasing reproducibility, the time until a measured value is eval-
uated as stable is longer.
4
With
<OK>
activate the
Stability control
function.
While the measured value does not meet the stability criterion,
the background color is red.
As soon as a stable measured value is recognized, the current
measurement data is downloaded to the interface. The mea-
sured values of all channels meeting the criterion for stability
control are marked by SC.
Note
You can prematurely terminate the
Stability control
function with
<OK>
manually at any time. If the
Stability control
function is termi-
nated prematurely, the current measurement data is not downloaded
to the interface.
5
Using
<SC>
or
<MODE>
, release the frozen measured value.
The [SC] status display disappears.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......