Conductivity
ProLab 4000
154
ba75646d03
12/2012
10.2.10 Selecting and determining the nonlinear temperature
compensation
TC nLin4
Preparatory activities
The nonlinear temperature compensation
TC nLin4
is determined from
several standards of unknown concentration.
You prepare the required standards yourself through equidistant dilu-
tion of the standard with the highest concentration e.g.:
Standard with the high-
est concentration
Volume share in %
Deionized water
Volume share in %
Standard 1
100
0
Standard 2
95
5
Standard 3
90
10
Standard 4
85
15
Standard 5
80
20
1
Highlight the channel of the conductivity measuring cell in the
measured value display.
2
If necessary, select the measured parameter with
<MODE>
.
3
Immerse the measuring cell in the test sample.
4
Open the input window for the measurement of the tempera-
ture dependency with
Calibration
/
TC calibration
/
TC nLin3
.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......