Technical Data
ProLab 4000
20
8
ba75646e03
12/2012
Cell constants
Cell constant C
Values
Can be calibrated in
the ranges
0.450 ... 0.500 cm
-1
0.585 ... 0.715
0.800 ... 1.200 cm
-1
Adjustable
0.010 cm
-1
(fixed)
0.090 ... 0.110 cm
-1
0.250 ... 25.000 cm
-1
Reference temperature
Reference tempera-
ture
Values
Adjustable
20 °C (Tr20)
25 °C (Tr25)
Accuracy
(± 1 digit)
Variable
Accuracy
Temperature of the test
sample
ϰ
/ Temperature compensation
None (Off)
± 0.5 %
Nonlinear (nLF)
± 0.5 %
0 °C ... + 35 °C
according to EN 27 888
± 0.5 %
+ 35 °C ... + 50 °C
Extended nLF function
according to SI Analytics
GmbH measurements
Linear (lin)
± 0.5 %
+ 10 °C ... + 75 °C
SAL
/ range
0.0 ... 42.0
± 0.1
+ 5 °C ... + 25 °C
± 0.2
+ 25 °C ... + 30 °C
TDS [mg/l]
± 1