ProLab 4000
pH
8
7
ba75646e03
12/2012
7.3.3
Calibration settings and calibration data
In the
Calibration
menu, you can
set the calibration interval,
define the calibration type and
view the calibration data.
Menu item
Setting
Description
Calibration
–
pH calibration type
–
AutoCal
(
)
Calibration with commercial buffer sets.
The buffer set is selected with the next
menu item (
Buffer sets
)
–
Buffer sets
DIN
(
)
TEC
...
Buffer sets to be used for pH calibration.
For details, see page 74.
–
VariCal
(
)
Calibration with any buffer solutions
–
Slope in
mV/pH
(
)
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.16 mV/pH ([determined slope/Nernst
slope ] x 100).
–
Zero point in
mV
(
)
pH
Unit for the zero point.
–
Calibration interval...
1 ... 999
Calibration interval...
for the pH electrode
(in days).
The meter reminds you to calibrate regu-
larly by the red frame around the CalClock
in the measuring window.
–
Calibration history...
Displays the calibration history of the last
calibrations.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......