12.5.3 Defining and activating limit values
With the limit values you define the measured values at which the curve
of the measurement developing is displayed gray if the measured are
below or above the limits.
Thus measured values below or above the limits can easily be recog-
nized.
1
Open the recorder menu,
Settings...Settings...
.
The
Recorder settings
window pops up.
2
Select the
Measured value
register to define the limit values for
the measured value.
3
Mark the selection field,
Switch on limit value monitoring
.
4
Enter the upper and lower limit value.
5
Select the
Temperature
register to define the limit values for
the measured value.
6
Mark the selection field,
Switch on limit value monitoring
.
7
Enter the upper and lower limit value.
8
Use
[OK]
to accept the new settings and close the dialog box,
Recorder settings
.
The new settings are active.
or
Select another register to make further settings.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......