Ion concentration
ProLab 4000
11
8
ba75646e03
12/2012
9.6.3
Double standard addition
In the
Double standard addition
procedure, a known amount of stan-
dard solution is added to the sample in two steps.
First, one percent of the sample volume is added, then 2 percent of the
sample volume is added.
The ion concentration in the sample is calculated from the change in
potential between the first and second addition of standard solution.
Measuring
1
Check the currently displayed ion.
If the ion to be measured is displayed incorrectly:
Select the ion type in the menu,
Measuring
/
ISE methods
/
Ion
type
.
2
In the
Measuring
/
ISE methods
menu, select the method,
Dou-
ble standard addition
.
The
ISE method
: The
Double standard addition
window pops
up.
The
Sample
register is open.
3
In the
Sample
register, enter the values for the sample volume
(
Sample volume
) and the volume of the ISA/TISAB solution
(
ISA/TISAB volume
) as necessary.
4
Use
[Dosing]
to automatically add the adjusted volume of the
ISA/TISAB solution (
ISA/TISAB volume
).
or
If the burette control is inactive:
Add the adjusted volume of the ISA/TISAB solution (
ISA/TISAB
volume
) manually.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......