Conductivity
ProLab 4000
15
8
ba75646d03
12/2012
Sample record:
Calibration evaluation
After the calibration, the meter automatically evaluates the current sta-
tus of the calibration. The evaluation appears on the display and in the
calibration record.
Print date
26.04.07 16:13
ProLab 4000 (06249876)
Current user: 1234567890
CALIBRATION Cond
Calibration date
20.04.07 10:14:03
LF413-ID (A062498765)
User: Administrator
Calibration interval:
150 days
Standard:
0,01 mol/l KCl
Cond./Tref 25:
1413 µS/cm
Cal. temp:
25,4 °C
Cell constant:
0.475 1/cm 25.0 °C
Sensor
+++
Note
If non-ID sensors are used, the sensor type and series number are not
output in the calibration record.
Display
Calibration record
Cell constant [cm
-1
]
+++
within the ranges
0.450 ... 0.500 cm
-1
0.585 ... 0.715 cm
-1
or
0.800 ... 1.200 cm
-1
Error
Error
outside the ranges
0.450 ... 0.500 cm
-1
or
0.800 ... 1.200 cm
-1
Eliminate the error according to chapter
16 W
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......